440Z Owner's Manual & Troubleshooting 

Z TROUBLESHOOTING & OPERATION NOTES:

 

Low Batteries & bad battery connections are the #1 cause of scale malfunction and inaccuracy

! We 

test all of our scale returns from consumers. Fully 60% of consumer returns are battery related 
problems. This sounds silly but it's true! A scale will perform slowly, or read inaccurately when it has low 
batteries. Please replace the batteries often (and only use good quality batteries). We include good 
quality batteries with all of our scales but batteries can run low in storage. If your scale simply won't 
turn on while on Battery power, this is often caused by loose battery connections. Battery prongs 
(terminals) are made of metal. They must be making good contact with your batteries in order for the 
scale to power on. You can use a paperclip to slightly bend the battery prongs to make them have a 
better connection. Also some poorly designed batteries have recessed or partially obstructed battery 
terminals. This may cause your prongs to be touching the plastic housing of the battery instead of the 
metal battery terminal.  

1) If the Display ever becomes locked showing "Out2", please recalibrate the scale as outlined in the 
manual  

2) If the Display ever becomes locked on LLLL or EEEE, this indicates that the scale was shocked, dropped 
or otherwise damaged and the delicate weighing sensors have been damaged. You can try recalibrating 
the scale (If the sensor has not been hurt too badly it will work again after recalibration). Otherwise you 
will have to follow the warranty instructions that came with your scale.  

3) If the display becomes locked on 8888, this often indicates low batteries. However sometimes it also 
may indicate a serious zero mark error. This means when you turn the scale on, it can't determine what 
zero is (a slight zero mark error will cause situation #1 above) Thus, if new batteries do not fix this error 
the scale will have to be sent to us for replacement under our Lifetime/30 year warranty program.  

4) If the display shows UNST, this means the scale is not stable. Please try operating the scale on a more 
stable surface and be sure nothing is on the tray or stuck under the tray when you turn it on. If this 
situation persists, it may be an indication of fatal load cell damage. Try calibrating the scale - if this does 
not work please follow the warranty instructions.  

Although the Z is designed to be extremely durable, It’s important that you never overload or 
drop/shock the scale. Scales are delicate instruments and unlike Cellular phones, scales have delicate 
sensors that determine how much an item weighs. If you drop or shock your scale, these sensors "feel" 
the shock and are sometimes destroyed. This happens with all digital scales. We design our scales to be 
as resistant to shock or drops as possible, however there is no way for us to protect 100% against load 
cell or sensor damage. A well-treated scale will provide years of reliable and accurate weighing. 
However an abused scale will only work until it’s sensors are damaged.  

   

1
2
3
4
5
Оглавление инструкции
  • Страница 1 из 6
    440Z Owner's Manual & Troubleshooting Z TROUBLESHOOTING & OPERATION NOTES: Low Batteries & bad battery connections are the #1 cause of scale malfunction and inaccuracy! We test all of our scale returns from consumers. Fully 60% of consumer returns are battery related problems. This sounds silly but
  • Страница 2 из 6
    PLEASE READ COMPLETE INSTRUCTIONS BEFORE USE Thank you for purchasing The Z. With normal care and proper treatment it will provide years of reliable service. Please read all operating instructions carefully. .KEEP THE FOLLOWING POINTS IN MIND: * If the display becomes locked on Out2 please
  • Страница 3 из 6
    c) Replace battery cover. DO NOT USE EXCESSIVE FORCE & DO NOT PRESS ON THE TRAY!!! 3) The scale is now ready for battery operation. (Advanced users only) CALIBRATION NOTE: There are two versions of the 120-Z scale. One has a switch on the bottom of the scale to change the mode from "Cal to Norm".
  • Страница 4 из 6
    1) Turn the scale OFF, Slide the switch on the bottom of the scale from "Norm" to "CAL". Then Place the scale on a Flat, very stable surface (Be certain the scale is OFF) 2) Press the key, a random number will appear on the display 3) Press the key again, the display will show 2Ero and then 100.0
  • Страница 5 из 6
    *Stable Indicator There is a stable indicator that appears on the left side of the LCD whenever the scale is stable and at Zero KEY PAD FUNCTIONS * UNIT (the left key) Unit (mode) selection: Press and hold the left key to change weight unit You can select g (grams), oz (ounces), dwt (pennyweight)