PLEASE READ COMPLETE INSTRUCTIONS BEFORE USE

  

Thank you for purchasing The Z. With normal care and proper treatment it will provide years of reliable 
service. Please read all operating instructions carefully.  

.

KEEP THE FOLLOWING POINTS IN MIND:

  

* If the display becomes locked on 

Out2

 please recalibrate the scale.  

* Avoid lengthy exposure to extreme heat or cold, your scale works best when operated at normal room 
temperature. 

Always allow the unit to acclimate to a normal room temperature for at least one hour 

before use.

  

* Allow sufficient warm up time. Turn the scale On and wait several seconds to give the internal 
components a chance to stabilize before weighing.  

* The cleaner the environment the better. Dust, dirt, moisture, vibration, air currents and proximity to 
other electronic equipment can all cause an adverse effect on the reliability and accuracy of your scale.  

* Handle with care. Gently apply all items to be weighed onto tray top. Although this scale is designed to 
be quite durable, try to avoid rough treatment as this may permanently damage the internal sensor and 
void your warranty.  

* Avoid shaking, dropping or otherwise shocking the scale. This is a precision instrument and MUST BE 
HANDLED WITH EXTREME CARE.  

*

IMPORTANT ADVICE: Place the item to be weighed on the platform, after the stable weight is 

displayed remove the item immediately. This will prolong the longevity and accuracy of this weighing 
instrument.

.  

.** These electronic scale are precision instruments. Do not operate near an in-use cell phone, cordless 
phone, radio, computer or other electronic device. These devices emit RF and can cause unstable scale 
readings. If your scale ever performs poorly, try moving the scale to a different room or location. This is 
a very precise scale - the display may seem to wander or jump when weighing. This is due to air currents 
or vibrations. Stable weighing is achieved when the display remains fixed for 3 seconds.  

OPERATION_____________________________________________ 

 

Battery Replacement:

  

1) Three #357 button batteries are required.  

2) To install batteries:  

a) Turn the scale upside-down and open the battery compartment  

b) Place batteries into battery compartment aligned correctly.  

1
2
3
4
5
Оглавление инструкции
  • Страница 1 из 6
    440Z Owner's Manual & Troubleshooting Z TROUBLESHOOTING & OPERATION NOTES: Low Batteries & bad battery connections are the #1 cause of scale malfunction and inaccuracy! We test all of our scale returns from consumers. Fully 60% of consumer returns are battery related problems. This sounds silly but
  • Страница 2 из 6
    PLEASE READ COMPLETE INSTRUCTIONS BEFORE USE Thank you for purchasing The Z. With normal care and proper treatment it will provide years of reliable service. Please read all operating instructions carefully. .KEEP THE FOLLOWING POINTS IN MIND: * If the display becomes locked on Out2 please
  • Страница 3 из 6
    c) Replace battery cover. DO NOT USE EXCESSIVE FORCE & DO NOT PRESS ON THE TRAY!!! 3) The scale is now ready for battery operation. (Advanced users only) CALIBRATION NOTE: There are two versions of the 120-Z scale. One has a switch on the bottom of the scale to change the mode from "Cal to Norm".
  • Страница 4 из 6
    1) Turn the scale OFF, Slide the switch on the bottom of the scale from "Norm" to "CAL". Then Place the scale on a Flat, very stable surface (Be certain the scale is OFF) 2) Press the key, a random number will appear on the display 3) Press the key again, the display will show 2Ero and then 100.0
  • Страница 5 из 6
    *Stable Indicator There is a stable indicator that appears on the left side of the LCD whenever the scale is stable and at Zero KEY PAD FUNCTIONS * UNIT (the left key) Unit (mode) selection: Press and hold the left key to change weight unit You can select g (grams), oz (ounces), dwt (pennyweight)
  • Страница 6 из 6