2

Connecting the Tubes

The tubes, CO

2

 gas cylinder and CO

2

 mixer are not included in the FV10i-LIV package. Please purchase

these accessories separately.

100% CO

2

 gas cylinder**

Tube

CO

2

 mixer*

Tube

Diameter converter
nipple

Tube
Inner dia. 2 x Outer
dia. 4 mm

Ultrapure water

Tube
Inner dia. 2 x Outer dia. 4 mm

Rear panel of
FV10C-W3

}As the CO

2

 concentration is reduced due to the incubator structure, please supply 6% CO

2

 gas. It will be turned into 5%

CO

2

 gas inside the incubator.

 *

Recommended CO

2

 mixer output flow: 150 ml/min. of 6% CO

2

 gas.

**

When a 6% CO

2

 gas cylinder is used, connect a flow meter in place of the CO

2

 mixer.