3. Do not subject the unit to excessive force, shock, dust, temperature, direct sunlight, heavy rain or 
humidity, which may result in malfunction, shorter electronic life span, battery damage or distorted 
parts.  

4. Do not tamper with the unit’s internal components. It will invalidate the warranty on the unit and 
may cause unnecessary battery damage and distorted parts.  

5. Place product on a flat stable surface before use.  

Simple Calibration 

1. Turn on the scale and wait for the “0g” to show on the display  

2. Put a 1000g standard weight on the platform and wait until it is stable.  

3. Press and hold the button for 3 seconds, the scale will be turned off temporary. Keep holding the 
button for another 20 seconds. The display will turn on again and show “1000”.  

The calibration is now completed.  

   

Remember that this is a precision instrument and must be handled with extreme care. If you overload 
your scale and "ERR" appears please try the advanced programming guide. Remember, 

Overload voids 

your warranty

 so please don't overload your scale!  

The 1scale is very easy to operate. However, there are certain caring and operation instructions you 
should read before your first use. We know taking the time to read these instructions is time consuming, 
but this scale is a precision instrument and must be handled with the upmost care. Only operate this 
scale in a clean, vibration free and air movement free envirornment. It is very sensitive!  

PROPER ENVIRONMENT AND OPERATION

 

 

Do not activate your scale near an active cell phone, cordless phone, or other electronic device. 
Doing so will cause your 1scale to temporarily malfunction and be extremely inaccurate.  

 

You must remove the yellow circular shipping spacer prior to installing batteries or turning on.  

 

Never turn the scale on with any objects already on the tray (clear the tray before turning on).  

 

Do not leave anything on the scale when not in use.  

 

If the scale ever "locks up" (reads 8888 for more then 15 seconds), simply remove and reinsert 
the batteries.  

 

Do not get the scale wet or hot and do not operate on any wet or hot surface.  

1
2
3
4
5
6
7
Оглавление инструкции
  • Страница 1 из 8
    1scale Owner's Manual Low Batteries, bad battery connections & Faulty AC Adaptors are the #1 cause of scale malfunction and inaccuracy! We test all of our scale returns from consumers. Fully 60% of consumer returns are battery related problems. This sounds silly but it's true! A scale will perform
  • Страница 2 из 8
    Note : If the battery symbol appears in the display during normal operation, it means low battery power. It is time to replace the battery. If the power is too low, it will show “Lo” when power on and then turn off automatically. Weighing Function 1. Turn the scale on by pressing the key below the
  • Страница 3 из 8
    3. Do not subject the unit to excessive force, shock, dust, temperature, direct sunlight, heavy rain or humidity, which may result in malfunction, shorter electronic life span, battery damage or distorted parts. 4. Do not tamper with the unit’s internal components. It will invalidate the warranty
  • Страница 4 из 8
     To clean the scale, wipe the surface with a clean damp cloth or mild non-abrasive cleaner. Do not use an abrasive cleaner it will damage the finish.  Do not overload the scale. That is, do not place anything that weighs more then 200 Grams on the scale. It may permanently damage the load cell. 
  • Страница 5 из 8
    6. The scale has an Auto-Off feature that turns off the scale after lapses of non-activity. If the scale is in a room with wind or air movement, it may sense the pressure of the air and not auto-off. In this case you must use the manual shutoff. Simply press the main big button to turn the scale
  • Страница 6 из 8
    9. Gently Place a 200g standard weight on the platform. The displayed figure will increase. 10. Wait 10 seconds for the display to stabilize and then gently press the Big Main Button once. The display should change to 0000. 11. Adjust the settings by using the Tare key (shifts the blinking digit)
  • Страница 7 из 8
    21. Turn the scale back on and test some weight readings.If the readings are still inaccurate, perform hard reset again from step 1 and try again. If after several attempts the scale still shows "err" or will not perform properly, please return it to us for repair. SPECIAL ADVANCED PROGRAMMING