Инструкция для MyWeigh 1SCALE

(скачивание инструкции бесплатно)
Формат файла: PDF
Доступность: Бесплатно как и все руководства на сайте. Без регистрации и SMS.
Дополнительно: Чтение инструкции онлайн
background image

1scale Owner's Manual 

Low Batteries, bad battery connections & Faulty AC Adaptors are the #1 cause of scale malfunction 
and inaccuracy

! We test all of our scale returns from consumers. Fully 60% of consumer returns are 

battery related problems. This sounds silly but it's true! A scale will perform slowly, or read inaccurately 
when it has low batteries. Please replace the batteries often (and only use good quality batteries). We 
include good quality batteries with all of our scales but batteries can run low in storage. If your scale 
simply won't turn on while on Battery power, this is often caused by loose battery connections. Battery 
prongs (terminals) are made of metal. They must be making good contact with your batteries in order 
for the scale to power on. You can use a paperclip to slightly bend the battery prongs to make them 
have a better connection. Also some poorly designed batteries have recessed or partially obstructed 
battery terminals. This may cause your prongs to be touching the plastic housing of the battery instead 
of the metal battery terminal. A Faulty AC adaptor can cause your scale to act unstable with numbers 
"jumping" all around. Please test your scale with a good set of batteries (instead of the AC adaptor) to 
determine if perhaps the AC adaptor is faulty.  

   

This page contains 

operating instructions for the 1Scale 200 and 1Scale 2000

 and 

Advanced 

Programming for the 1Scale 200

. Please make sure you are looking at the correct manual for your 

model scale.  

   

1SCALE 2000 OPERATING INSTRUCTIONS 

The 1scale is very easy to operate. However, there are certain caring and operation instructions you 
should read before your first use We know taking the time to read these instructions is time consuming, 
but this scale is a precision instrument and must be handled with the utmost care.  

Technical Specification 

Maximum Capacity : 4lbs. 6½oz / 2000 grams  

Resolution : 1/8 oz / 1 gram  

Power : 4 AAA Batteries  

Battery Installation 

1. Press and push open the battery cover located at the bottom of the unit.  

2. Insert 4 AAA size batteries matching the (+) and (-) polarity covers with the battery. Close the battery 
cover until it clicks shut.  

1
2
3
4
5
6
7
Оглавление инструкции
  • Страница 1 из 8
    1scale Owner's Manual Low Batteries, bad battery connections & Faulty AC Adaptors are the #1 cause of scale malfunction and inaccuracy! We test all of our scale returns from consumers. Fully 60% of consumer returns are battery related problems. This sounds silly but it's true! A scale will perform
  • Страница 2 из 8
    Note : If the battery symbol appears in the display during normal operation, it means low battery power. It is time to replace the battery. If the power is too low, it will show “Lo” when power on and then turn off automatically. Weighing Function 1. Turn the scale on by pressing the key below the
  • Страница 3 из 8
    3. Do not subject the unit to excessive force, shock, dust, temperature, direct sunlight, heavy rain or humidity, which may result in malfunction, shorter electronic life span, battery damage or distorted parts. 4. Do not tamper with the unit’s internal components. It will invalidate the warranty
  • Страница 4 из 8
     To clean the scale, wipe the surface with a clean damp cloth or mild non-abrasive cleaner. Do not use an abrasive cleaner it will damage the finish.  Do not overload the scale. That is, do not place anything that weighs more then 200 Grams on the scale. It may permanently damage the load cell. 
  • Страница 5 из 8
    6. The scale has an Auto-Off feature that turns off the scale after lapses of non-activity. If the scale is in a room with wind or air movement, it may sense the pressure of the air and not auto-off. In this case you must use the manual shutoff. Simply press the main big button to turn the scale
  • Страница 6 из 8
    9. Gently Place a 200g standard weight on the platform. The displayed figure will increase. 10. Wait 10 seconds for the display to stabilize and then gently press the Big Main Button once. The display should change to 0000. 11. Adjust the settings by using the Tare key (shifts the blinking digit)
  • Страница 7 из 8
    21. Turn the scale back on and test some weight readings.If the readings are still inaccurate, perform hard reset again from step 1 and try again. If after several attempts the scale still shows "err" or will not perform properly, please return it to us for repair. SPECIAL ADVANCED PROGRAMMING
  • Страница 8 из 8